VS 8/VS 8A Vital Signs Monitor Operator’s Manual 16 - 27
4. Set Multicast Address.
16.15.11Setting the Network Service Quality Level
To set the quality of service (QoS), follow this procedure:
1. Select the Main Menu quick key → from the System column select Maintenance 
→ input the required password → select .
2. Select the Network Setup tab → QoS tab.
3. Select QoS Level For Realtime Monitoring to set the service quality of network 
connection for realtime monitoring, for example parameter measurements and 
waveforms, alarms, etc.
4. Set QoS Level For Others to set the service quality of network connection for 
non-realtime monitoring, for example history data, printing, etc.
16.15.12Setting MLDAP
To access the MLDAP server, set your monitor as follows: 
1. Select the Main Menu quick key → from the System column select Maintenance 
→ input the required password → select .
2. Select the Network Setup tab→ MLDAP tab.
3. Set IP address and port of the MLDAP server.
16.15.13Testing MLDAP Server Connection
To test whether the monitor is properly connected with the MLDAP server, follow this 
procedure: 
1. Select the Main Menu quick key → from the System column select Maintenance 
→ input the required password → select .
2. Select the Network Setup tab→ MLDAP tab.
3. Select Network Test.
16.15.14Setting VitalsLink
To set VitalsLink, follow this procedure:
1. Select the Main Menu quick key → from the System column select Maintenance 
→ input the required password → select .
2. Select the Network Setup tab → VitalsLink tab.
3. Switch on Enable.
4. Set Automatically Logout after as needed.
5. From the General Connectivity area, set the Master Server Address, Context 
Root, and Organization ID. Enable Local CSI as needed.
16 - 28 VS 8/VS 8A Vital Signs Monitor Operator’s Manual
6. From the Security area, set the Auth Type, User Name, User Password, Te na nt  
Type, and Transport Protocol.
7. Set Data Encoding Type as needed.
16.16 Enabling Bluetooth
To enable the Bluetooth function, follow this procedure:
1. Select the Main Menu quick key → from the System column select Maintenance 
→ input the required password → select .
2. Select the Bluetooth.
3. Switch on Enable.
4. Set Auto Unpair On Disconnection as needed.
16.17 Defining Other Functions
16.17.1 Setting the SpO
2
 Tone Mode
The monitor adjusts the QRS tone (pitch tone) according to the SpO
2
 values. To set the 
SpO
2
 tone mode, follow this procedure:
1. Select the Main Menu quick key → from the System column select Maintenance 
→ input the required password → select .
2. Select the Other tab. 
3. Set SpO2 Tone to Mode 1 or Mode 2. The pitch changes more rapidly in Mode 2 
than in Mode 1, as saturation levels change.
The SpO
2
 tone mode chart for SpO
2
 between 40% to 100% is as follows: