7. As required, replace the seal for pressure sampling connector (M6M-010063---), seal for
fresh gas and ACGO (M6M-010006---), seal for drive gas and APL discharge
(M6M-010058---) every 12 months.
8. As required, replace the seal for bellows housing (0601-20-78848) and folding bag
(0601-10-69901) every 12 months.
Seal(M6M-010058---)
Seal(M6M-010063---)
Seal(M6M-010006---)
Folding bag
(0601-10-69901)
Seal for bellows
housing
(0601-20-78848)