4-7
9. As required, replace the seal for axis of bag/mechanical ventilation switch
(0030-10-13077) every 12 months. For details, refer to 6Repair and Disassembly.
Screws to be unscrewed
Pull out the pin axis after
removing the seal herein
Pin axis pulled out
Seal0030-10-13077