Safety Precautions 1-3
1.4 Safety Precautions
Please observe the following precautions to ensure patient and operator’s safety when using this
system.
DANGER:
DO NOT use flammable gasses, such as anesthetic gas or
hydrogen, or flammable liquids such as ethanol, near this system,
because there is danger of explosion.
WARNING:
1.
Do connect the power plug of this system and power plugs of
the peripherals to wall receptacles that meet the ratings
indicated on the rating nameplate. Using a multifunctional
receptacle may affect the system protective grounding
performance, and cause the leakage current to exceed safety
requirements.
Use the cable provided with this system to connect the printer.
Other cables may result in electric shock.
You must use the power adapter provided with the system;
otherwise electric shock may result.
You can only adopt the power supply method provided by
Mindray, other power supply modes (e.g. using a UPS) may
result in electric shock.
2.
Connect the protective grounding conductor before turning
ON the system. Disconnect the grounding cable after turning
OFF the system. Otherwise, electric shock may result.
3.
For the connection of power and grounding, follow the
appropriate procedures described in this operators manual.
Otherwise, there is risk of electric shock. DO NOT connect the
grounding cable to a gas pipe or water pipe; otherwise,
improper grounding may result or a gas explosion may occur.
4.
Before cleaning the system, disconnect the power cord from
the outlet. System failure and electric shock may result.
5.
This system is not water-proof designed. Do Not use this
system in any place where water or any liquid leakage may
occur. If any water is sprayed on or into the system, electric
shock
or device malfunction may result. If water is accidentally
sprayed on or into the system, contact Mindray Customer
Service Department or sales representative.
6.
Do not use a transducer that has a damaged, scratched
surface, or exposed wiring of any kind. Immediately stop
using the transducer and contact Mindray Customer Service
Department or sales representative. There is risk of electric
shock if using a damaged or scratched transducer.
7.
DO NOT allow the patient to contact the live parts of the
ultrasound system or other devices, e.g. signal I / O ports.
Electric shock may occur.
8.
Do not use an aftermarket probe other than those specified
by Mindray. The probes may damage the system causing a
profound failure, e.g. a fire in the worst case.