13-12 Probes and Biopsy
NOTE: 1. Observe the graph here carefully to immerse the transducer. Only soak parts
of the transducer below the strain relief.
2. Repeated disinfection will eventually damage the probe, please check the
probe performance periodically.
Compatible Disinfectants
Please see Mindray Transducer Disinfectant Recommendation for the disinfectant use in
details.
13.1.6 Environment
Working condition
Use the probes in the following ambient conditions:
Probe
Ambient
temperature
Relative humidity
Atmospheric
pressure
3C5P
0℃~40
30%85%RHno
condensation
700hPa1060hPa
6C2P
0-40
30%~85%RHno
condensation
700 hPa ~ 1060 hPa
6CV1P
0℃~40
25%90%RHno
condensation
550hPa1060hPa
CB10-4P
0℃~40
30%85%RHno
condensation
700hPa1060hPa
7L4P
0℃~40
30%85%RHno
condensation
700hPa1060hPa
7L5P
0℃~40
30%85%RHno
condensation
700hPa1060hPa
L14-6P
0℃~40
30%85%RHno
condensation
700hPa1060hPa
Strain relief
Probe handle