Safety Precautions 1-5
3.
Precautions concerning movement of the system:
When you place the system on the mobile trolley and move
them together, you must secure all objects on the mobile
trolley to prevent them from falling. Otherwise you should
separate the system from the mobile trolley and move them
individually.
When you have to move the system with the mobile trolley
upward or downward the stairs, you must separate them first
and then move them individually.
Object placed on the monitor may fall and injure an
individual when moving.
Fasten and fully secure any peripheral device before moving
the system. A loose peripheral device may fall and injure an
individual.
4.
DO NOT expose the system to excessive vibration through
transportation. Mechanical damage may result.
5.
Do not connect this system to outlets with the same circuit
breakers and fuses that control the current of devices such
as life-support systems. If this system malfunctions and
generates overcurrent, or when there is an instantaneous
current at power ON, the circuit breakers and fuses of the
building
s supply circuit may be tripped.
6.
Always keep the system dry. Avoid transporting this system
quickly from a cold place to a warm place; otherwise
condensation or water droplets may form allowing a short
circuit and possible electric shock.
7.
If the circuit protector is tripped, it indicates that the system
or a peripheral device was improperly shut down and the
system is unstable. You cannot repair the system under this
circumstance and must call the Mindray Customer Service
Department or sales representative.
8.
There is no risk of high-temperature burns during normal
ultrasound examinations. It is possible for the surface
temperature of the transducer to exceed the body
temperature of a patient due to environmental temperature
and exam type combinations. Do not apply the transducer to
the same region on the patient for a long time. Apply the
transducer only for a period of time required for the purpose
of diagnosis.
9.
The system and its accessories are not disinfected or
sterilized prior to delivery. The operator is responsible for the
cleaning and disinfection of transducers and sterilization of
biopsy brackets according to the manuals, prior to the use.
All items must be thoroughly processed to completely
remove harmful residual chemicals, which will not only
harmful to the human body, but also damage the accessory.
10.
It is necessary to press [End Exam] to end the current scan
that is in progress and clear the current Patient Information
field. Otherwise, new patient data may be combined with the
previous patient data.