13-14 Probes and Biopsy
Probe
Ambient
temperature
Relative humidity
Atmospheric
pressure
7L5P
-20℃~55℃
30%~95%RH(no
condensation)
700 hPa ~ 1060 hPa
L14-6P
-20℃~55℃
30%~95%RH(no
condensation)
700 hPa ~ 1060 hPa
V10-4BP
-20℃~55℃
30%~95%RH(no
condensation)
700 hPa ~ 1060 hPa
7LT4P
-20℃~60℃
10%~95%RH(no
condensation)
500 hPa ~ 1060 hPa
6LE7P
-20℃~55℃
30%~95%RH(no
condensation)
700 hPa ~ 1060 hPa
2P2P
-20℃~55℃
30%~95%RH(no
condensation)
700 hPa ~ 1060 hPa
P7-3P
-20℃~55℃
30%~95%RH(no
condensation)
700 hPa ~ 1060 hPa
35C50P
-20℃~55℃
30%~95%RH(no
condensation)
700 hPa ~ 1060 hPa
75L38P
-20℃~55℃
30%~95%RH(no
condensation)
700 hPa ~ 1060 hPa
D6-2P
-10℃~50℃
20%~95%RH(no
condensation)
700 hPa ~ 1060 hPa
C6-2P
-20℃~55℃
20%~95%RH(no
condensation)
700 hPa ~ 1060 hPa
7L4BP
-20℃~55℃
20%~95%RH(no
condensation)
700 hPa ~ 1060 hPa
1. To prevent the probe from being damaged, DO NOT store it where it may be exposed to:
Direct sunlight or X-rays
Sudden changes in temperature
Dust
Excessive vibration
Heat generators
2. When the probe is sent to the MINDRAY Customer Service Department or a sales
representative for repair, be sure to disinfect it and keep it in the carrying case to prevent
infection.
3. Sterilize the carrying case as necessary.