1-8 Safety Precautions
6.
Never immerse the probe connector into liquids such as
water or disinfectant because
the connector is not
waterproof. Immersion may cause electric shock or
malfunction.
7.
A probe sheath must be installed
over the probe before
performing intra-cavity, biopsy or intra-operative
examination.
CAUTION:
1. When using the probe, wear sterile gloves to prevent infection.
2.
Be sure to use ultrasound gel. Please
use the ultrasound gel
compliant with the relevant local regulations. And manage the
ultrasound gel properly to ensure that it does not become a
source of infection.
3.
In normal diagnostic ultrasound mode, there is no danger of a
normal-
temperature burn; however, keeping the probe on the
same region
of the patient for a long time may cause such a
burn.
4.
Do not use the carrying case for storing the transducer. If the
carrying case is used for storage, it may become a source of
infection.
5.
It is required to practice ALARA when operating ultrasound
system. Minimize
the acoustic power without compromising
the quality of images.
6.
The probe and accessories supplied with it are not delivered
disinfected or sterilized. Sterilization (or high-
level disinfect)
before use is required.
7.
Disposable components are packaged sterile and are single-
use only. Do
not use if integrity of packaging violated or if
expiration date has passed. Please
use the disposable
components compliant with the relevant local regulations.
8.
Please use the disinfection o
r sterilization solution that
recommended in this operators manual, otherwise Mindray
will not be liable for damage caused by other solutions. If you
have any questions, please contact Mindray Customer Service
Department or sales representative.
9.
The
probe sheath contains natural rubber that can cause
allergic reactions in some individuals.
10.
Do not use pre-lubricated condoms as a sheath. Lubricant
may not be compatible with the transducer material and
damage may result.