EasyManua.ls Logo

Mitutoyo ABSOLUTE ABS AT1123 - Safety Precautions; Electromagnetic Compatibility (EMC) Precautions; Required Installation Environment

Mitutoyo ABSOLUTE ABS AT1123
63 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ii
No. 99MBE095B
Safety Precautions
Observe the following descriptions to make full use of the performance of this product:
CAUTION
Read this User's Manual thoroughly before operating the instrument.
Before connecting this product to the main unit of the system, make sure that the power for the control unit is
turned o󰀨.
To maintain the shielding e󰀨ect, tighten rmly the screws of the connectors of each connecting cable. Also,
to prevent defective contacts, do not touch the connecting terminals of the connectors with bare hands.
Electromagnetic Compatibility (EMC)
This product complies with the EMC Directive. Note that in environments where electromagnetic
interference exceeds EMC requirements dened in this directive, appropriate countermeasures are
required to ensure product performance.
EMC Directive EN61326-1
Immunity test requirement: Clause6.2 Table 2
Emission limit: Class B
Authorized representative and importer in the EU:
Mitutoyo Europe GmbH
Borsigstrasse 8-10,41469 Neuss,Germany
For the EU Directive
Required Environment for Installation
Vibration
To install this product unit in a machine, select a location where there is as little vibration as possible.
If this product is used for an extended period of time in a location where there is a substantial amount
of vibration, the built-in precision parts may be damaged, thereby adversely inuencing the perfor-
mance of this product.
Shock, dust, water/oil protection
To protect the scale unit from being directly exposed to machining oil and chips, or from being
bumped by a workpiece, etc., prepare a cover that protects the entire scale unit.
Ambient temperature and humidity
This product should be operated in an environment where the temperature is between 0°C and 50 °C
and where the relative humidity is between 20 and 80 % RH. Do not use this product in a place where
sudden changes in temperature or humidity are observed.

Table of Contents

Related product manuals