58
5 Setting Parameters
No. 99MAH057A
(Short press) / (Long press)
Tips
When [Auto] is selected for the analog bar scale, it automatically switches to the
analog bar scale when the conditions below are present.
• When peak detection runout width (TIR) displayed:
The runout width is the analog bar display entered within the analog bar display
range
• When tolerance judgment function ON:
The preset value is the analog bar display entered in the analog bar display range
• When the resolution switched:
Analog bar display that is identical to the resolution