7 Error Displays and Countermeasures
7
Error Displays and Countermeasures
No. Error name LCD display Cause Countermeasures
-
ABS Synthesis
Error
Last digit is E.
The sensor
signal cannot be
synthesized.
Although this may occur
during high-speed
plunger movement,
thereisnoe󰀨ecton
measurement.
Use the product as is.
* If it occurs while the
plunger is not moving,
the sensor may have
failed. Please contact
the agent where you
purchased the product
or a Mitutoyo sales/
service representative.
( "11O󰀨-SiteRepairs
(Subject to Charge)" on
page 93)
-
Battery voltage
decrease display
Battery voltage is
low.
Replace with a new
battery.
15
Low Battery
Voltage Error
Err 15
Cannot perform
measurement
because the
battery is low.
Replace with a new
battery.