RS-232C / RS-LINK
FUNCTION
This chapter describes the RS-232C communication function.
6
6.1 RS-232C Communication Function
6.1.1 List of commands
Command format Response output Operation content
GA**CRLF G#**,+01234.567CRLF A display value is outputted via RS-232C.
“**" indicates gage channel numbers 01 to 99 (all channels for 00).
If LINK connection does not need to be made, set the channel number to 01.
“#” indicates the type of data (N: current value, X: maximum value, M:
minimum value, and W: TIR).
CRLF stands for carriage return (CR) and line feed (LF).
CN**CRLF CH**CRLF The display is switched to the current value.
CX**CRLF CH**CRLF The display is switched to the maximum value.
CM**CRLF CH**CRLF The display is switched to the minimum value.
CW**CRLF CH**CRLF The display is switched to the TIR value.
CR**CRLF CH**CRLF The display is zero-set.
CL**CRLF CH**CRLF The peak value is cleared.
CP**,+01234567CRLF CH**CRLF The preset value is inputted.
Input a preset value and a tolerance limit with a sign and a numeric value of
8 digits without appending a decimal point.
CD**,+01234567CRLF CH**CRLF Input tolerance limit S1.
Perform tolerance setup in the order of CD and CG for 3-step tolerance
judgment, and in the order of CD, CE, CF, and CG for 5-step tolerance
judgment.
An error message is outputted if there is a difference in tolerance limit order,
or in the number of steps between the setting and data to be sent, or if
incorrect data exists.
If this is the case, repeat setup from the beginning of the CD command.
CE**,+01234567CRLF CH**CRLF Input tolerance limit S2.
CF**,+01234567CRLF CH**CRLF Input tolerance limit S3.
CG**,+01234567CRLF CH**CRLF Input tolerance limit S4.
CS**CRLF CH**CRLF An error is canceled.
VS**,+$CRLF CH**CRLF Spindle control
Sign +: Moves up the spindle., -: Moves down the spindle.
$: Speed specification 0: Stop, 1: 2mm/s, 2: 4mm/s, 3: 8mm/s approx.
VT**CRLF CH**#CRLF Status of spindle condition
In place of #, 0: Normal, 1: Upper dead point limit, 2: WORK ON
Channel number 00 cannot be used.
*The output at an error is “Error$$CRLF” (For $$: Error code, see “5. Error Display”).
Note 1: After receiving a response output for the command, send the next command. If there is no response to the
command, clear the communication buffer, and then resend the command after 1 second or more.
Note 2: The RS communication function is paused during key operation (parameter, preset value, or tolerance limit
setup), and the command and data output is executed when the counter returns to the count mode.
Note 3: Clear the count standby mode with CS00CRLF (all channel specification).
Note 4: It takes about 100 ms for the command processing time (When smoothing is “soft”)
Note 5: If the peak mode is switched in the RS-232C command, memory back up cannot be performed.
No. 99MBC082B
6 - 1