ALTAIR 4X
GB
47
(5) Reinstall the case front.
(6) Re-install the screws.
(7) Turn on the device.
(8) Calibrate the device after the sensors have stabilised.
If the sensor replaced is the same as
the previous sensor:
If the sensor replaced is not the same
as the previous sensor or this sensor
channel was deactivated:
- The device starts up normally.
- The device automatically senses
that a new sensor is installed and
displays the "SENSOR DSCVRY"
screen.
- The device automatically senses
the difference and displays
"SENSOR CHANGE".
-"ACCEPT?" appears on the display.
Accept the change with
button or reject with
button.
Go into the sensor setup and
turn on the appropriate sensor
( chapter 3.4).
Danger!
Calibration is required after a sensor is installed; otherwise, the device
will not perform as expected and persons relying on this product for their
safety could sustain serious personal injury or death.
Allow sensors to stabilise at room temperature for at least 30 minutes
before calibration (chapter 3.9).