3.4 Device Setup
The setup menus can be accessed only when the device is turned on while pressing
and holding the button.
This mode can only be activated at device turn-on.
The operation is as follows:
(1) Press and hold the button while turning the device on.
Use the and buttons to enter the setup password.
The default password is "672".
(2) Press ON/OFF button to enter the setup menus.
Incorrect password: device enters the Measure mode.
Correct password: device continues/beeps three times.
In the Setup mode:
- Press the ON/OFF button to store chosen value or go to the next page.
- Press button to increase values by one or toggle a selection on or off.
- Press and hold button to increase values by 10.
- Press button to decrease values by one or toggle a selection on or off.
- Press and hold button to decrease values by 10.
The following options are available by pressing the and buttons:
- Sensor Setup [SENSOR SETUP]
- Calibration Setup [CAL SETUP]
- Alarm Setup [ALARM SETUP]
- Setup Time and Date [TIME SET]
- EXIT
The password can be changed through the MSA Link software.
FASIT Srl - Via Orsera 13/b - 34145 TRIESTE - ITALY - Tel +39 040 825604 - www.fasit.it