EasyManua.ls Logo

MSA ALTAIR 5X PID - Page 37

MSA ALTAIR 5X PID
91 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ALTAIR 5X PID
37
Operation
US
(3) Set values for Sensor Alarm by pressing the T or S button.
(4) Press the q button to confirm set value.
(5) Repeat setting for all other sensors.
(6) Press the S button to return to the Alarm Options menu.
(7) Repeat setting for all other alarm types.
Instrument Options
The Instrument Options menu allows modification of different device options:
- Sensor Setup (enable/disable the channel)
- Language Setup
- Time Date Setup
- Datalog Intervals
- Stealth Mode
- Operating Beep
- Backlight Options
- VOC Gas Setup
- Bluetooth
Press
- the T button go to next page
the S button to go previous page
the q button to enter setup.
Setting Sensor Options
(1) Press the q button to enter setup.
Z Following screen displays:
Fig. 10 Sensor Options Setup
(2) Press the T button to select sensor, press the q button to make changes.
Z The sensor information is displayed and the sensor can be enabled or disabled.
(3) Change status by pressing the T or S button.
SETTINGS
Other operations such as changing the combustible gas type (Methane, Butane,
Propane etc.) and units (ppm to mg/m3) are only possible using the MSA Link software.

Table of Contents