Appendix: Instrument SpecificationsMSA
ULTIMA X
®
Series
GB
81
1)
See Calibrator/Controller manual for ALERT option.
2)
SWAP delay signal is 60 s if enabled, otherwise 0 s.
3)
See Calibrator/Controller manual for SWAP delay options.
4)
Alarm operation will follow if the alarms are enabled.
Sensor
missing /
Countdown
OFF ON 3.0 mA if SWAP delay
time-out expired
2)
,
SWAP delay disabled
or FAULT
De-energised if SWAP
delay timeout
2)
expired,
SWAP delay
3)
disabled
or FAULT
Last gas value, if
SWAP delay enabled
and SWAP delay
3)
timeout not expired
Energised, if SWAP delay
3)
enabled and SWAP de-
lay timeout has not ex-
pired
2)
.
Sensor
calibration
OFF ON 3.75 mA if calibration
signal enabled and
ALERT option
1)
ena-
bled; gas value if cali-
bration signal is
disabled
Energised if ALERT op-
tion disabled
21.0 mA for 02 if cali-
bra-tion signal ena-
bled and ALERT
option
1)
disabled
De-energised if ALERT
option enabled
1)
Calibration
4-20
OFF ON 4 mA, if 4 mA calibra-
tion selected
Energised if ALERT op-
tion disabled
20 mA, if 20 mA
calibration selected
De-energised if ALERT
option enabled
1)
Calibration
fault
OFF ON Gas value De-energised two
seconds every minute
Under-
range
OFF ON 3.0 mA De-energised
Overrange/
LOC
ON
4
OFF
4
21 mA Energised
Operating
mode
LED
4 – 20 mA Fault relay
GREEN RED

Table of Contents

Other manuals for MSA Ultima X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the MSA Ultima X Series and is the answer not in the manual?

MSA Ultima X Series Specifications

General IconGeneral
BrandMSA
ModelUltima X Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals