Page #55 background imageLoading...
Page #55 background image
Technical DataMSA
ULTIMA X
®
Series
GB
55
(4) The ULTIMA X
®
Series Gas Monitor is shipped with the Sensor Swap Delay
enabled. This means that the 4-20 mA output signal and the FAULT relay will
hold off a fault indication for 60 seconds after the missing sensor indication is
displayed on the instrument. This setting allows the operator to exchange sen-
sor modules without a FAULT indication.
(5) Refer to Chapter "Calibration".
7 Technical Data
7.1 Dimensions, Weight
Alarm setpoints and relay functions [energised/de-energised] will not
change when changing a sensor from its current type to the same gas
type.
Alarm setpoints and the upscale/downscale relay function will change to
the new sensor’ settings when changing a sensor module from its cur-
rent type to a different gas type.
Gas types Combustible gases, oxygen and toxic gases
XE Dimensions W x H x D [mm]: 162 x 262 x 100
Weight approx. 5 kg
XIR Dimensions W x H x D [mm]: 320 x 150 x 100
Weight approx. 4.75 kg
XA Dimensions W x H x D [mm]: 130 x 240 x 76
Weight approx. 0.7 kg

Table of Contents

Other manuals for MSA Ultima X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the MSA Ultima X Series and is the answer not in the manual?

MSA Ultima X Series Specifications

General IconGeneral
BrandMSA
ModelUltima X Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals