EasyManua.ls Logo

multivac P200 - Page 76

multivac P200
88 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Troubleshooting
Faults without diagnostic message
76 03.06.2014
Symptom Cause Solution
Ambient temperature too low. Operate the machine only at
permissible ambient tempera-
ture, see "Technical Specifica-
tions".
Vacuum pump defective. Notify the service.
Chamber lid pressed closed
too briefly.
Press the chamber lid shut
more evenly, more firmly and
for longer.
Chamber lid gasket is dirty. Clean the chamber lid gasket.
Chamber lid does not
remain closed despite
pump running.
Chamber lid gasket is dam-
aged.
Change chamber lid gasket.
The vacuum pump only
runs for a very short
period.
Evacuation time is set incor-
rectly.
Increase evacuation time.
Evacuation time is set incor-
rectly.
Increase evacuation time.
Gas flushing time is set incor-
rectly.
Reduce the gas flushing time.
Pouch neck is clamped by
chamber lid.
Insert the film pouch so that
the pouch neck lies within the
chamber. Cut off the pouch
neck if necessary.
Chamber lid gasket is dirty. Clean the chamber lid gasket.
Insufficient oil quantity or oil in
the vacuum pump is too old.
Refill or replace oil.
Chamber lid gasket is dam-
aged.
Change chamber lid gasket.
Moist or liquid products give
off steam during the evacua-
tion process.
Abort the evacuation process
with the <Stop> key in case
the chamber lid mists up.
Poor vacuum in the
pack.
Evacuation system is leaking. Notify the service.
Sealing time is set incorrectly. Increase the sealing time in
small increments.
Pouch neck is clamped by
chamber lid.
Insert the film pouch so that
the pouch neck lies within the
chamber. Cut off the pouch
neck if necessary.
Pack is leaking.
Seal seam is dirty. Keep the inside of the film
pouch in the area of the seal
seam clean.

Table of Contents

Related product manuals