31
[3] Apply Control Solution
Shake the control solution bottle.
Discard the  rst drop of control
solution.
Wipe the bottle tip clean using a clean
tissue or paper towel.
Dispense a second drop onto a clean surface, such as an unused
re-sealable plastic bag.
Bring the tip of the test strip to the control solution sample.
The MyStar Extra® Meter will automatically detect if a test is performed
with BGStar® Control Solution.
This is the visual  ll
window. Do not put
control solution here.
APPLY
CONTROL
SOLUTION
HERE.
Performing a Control Solution Test