EasyManua.ls Logo

nal von minden Reactif Touch - Page 32

nal von minden Reactif Touch
88 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
28
nal von minden GmbH · Manual
Reac f Touch
The third two-beep sounds prompt the user to dip the third strip into the specimen. The
analyzer will complete the fi rst test and print the result if the analyzer has been enabled
to print automa cally. At this  me, the user should dip the third strip into the specimen,
lay the strip on the work bench, and put the second strip onto the test table. 15 seconds
a er the third two-beep sounds, the analyzer will start to run the test with the second
test strip.
Cau on: It is important to lay the strip on a fl at surface for about 30 seconds. Do not
put the strip on the test table immediately a er dipping since it may cause
false results.
Note: If the analyzer has been set up to fl ag a microscopy test for one or more para-
meters (refer to Sec on 7 Analyzer Setup: Test Se ngs: Abnormal Se ngs:
Flags for Microscopy), the results meet the Microscopy se ng value (Refer
to Sec on 7 Analyzer Setup: Test Se ngs: Abnormal Se ngs: Microscopy
se ngs) is also displayed on a Microscopy Report screen.