64
nal von minden GmbH · Manual
Reac f Touch
Sec on 11 Troubleshoo ng
General Informa on
The Reac f Touch Urine Analyzer will operate properly when direc ons for use and pro-
per cleaning of the instrument are followed.
Error Messages
Error messages will be displayed when the Reac f Touch Urine Analyzer detects anything
which needs a en on. The format of this advisory informa on depends upon the im-
portance of the problem and the mode in which the instrument is being used.
Note: To correct an error, see the List of Errors and Advisory Messages located
at the end of this sec on.
Errors which Disable the Instrument
If the error prevents the instrument from being used, all selec on areas on the screen
will be disabled. Taking correc ve ac on will remove the error alert screen and will
allow use of the instrument again.
Other Errors
There are certain errors which need to be corrected to enable tes ng of specimens, but
do not prevent other instrument func ons from being used. The user must carry out the
correc ve ac on to enable tes ng.
Advisory Messages
Errors of less importance will be presented via a message on the ini al screen. A er
taking correc ve ac on, the message will be removed from the display. If more than one
of this class of errors occurs, clearing one message will enable the next to be displayed in
order of importance to a user.
Results Alert
If an error occurs during tes ng and the test cannot con nue because of the error, the
error will be presented via the Results Alert screen. This screen will provide details of
the error and show that the test has been cancelled. The test table will be extended so
that the urinalysis strip can be removed.
Ba ery Power Icon
The ba ery icon indicates the power level of the ba ery. When the ba ery power
level is reduced, the analyzer will s ll con nue to test with only an advisory message
displayed on the ini al screen. If the ba ery level falls too low to power the analyzer, all