External case cleaning
1. Turn the analyzer Off by pressing the On/Off bu on for 2 seconds.
2. Wipe the outside (including the display) with a damp (not wet) cloth and a mild
detergent.
Cau on: Do not use any type of solvent, oil, grease, silicone spray, or lubrica on on
the analyzer.
Do not spray glass cleaner directly onto the screen.
Do not use laboratory wipes, such as Kimwipes, since they may scratch the screen.
Care should be taken to avoid liquid from entering the printer compartment.
Note: The display may be disinfected using the same solu ons as the test table
(see earlier in this sec on). Wipe the solu on on and leave on for 10 minutes.
Wipe clean using a clean cloth dampened with water, then dry.
Changing Ba eries
The Reac f Touch Urine Analyzer is designed to run the maximum number of tests
(approximately 100 with the printer OFF) from a set of ba eries. To achieve this, the
Power Save feature is always ac vated when the instrument is powered by ba eries.
If the instrument is not used for 3 minutes when ba ery-powered, it will automa cally
power down.
A ba ery power icon is shown in the bo om right corner of the ini al screen when the
analyzer is being powered by ba eries. The number of segments displayed represents
the amount of power remaining in the ba eries.
When the power is reduced, but tes ng can con nue, an advisory message will be
displayed on the ini al screen. The analyzer will stop tes ng a er only 10 strips when
the power is low.
Cau on: If you do not change the ba eries and the level becomes too low to power
the analyzer, the error will become cri cal and all selec on areas on the
screen will be disabled un l the ba eries are replaced.