EasyManua.ls Logo

nal von minden Reactif Touch - Page 64

nal von minden Reactif Touch
88 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
60
nal von minden GmbH · Manual
Reac f Touch
4. If the white calibra on bar is clean and unmarked, replace the table into the ana-
lyzer by holding the table at the end opposite the white calibra on bar, with the
white calibra on bar facing upwards. Push the test table fi rmly but slowly just over
half way into the analyzer. Replace the test table insert.
5. If the white calibra on bar is dirty or discolored, gently wipe and clean it with a
new co on swab or lint-free cloth that is wet with dis lled water.
Cau on: Care should be taken not to scratch the white calibra on bar. Solvents of
any kind must not be used to clean the bar.
6. Allow the calibra on bar to air dry and then inspect the surface for dust, foreign
material, scratches, or scuff s. If the calibra on bar cannot be cleaned or is s ll mar-
ked, obtain a new test table.
7. Reinsert the test table as described previously in step 4.