EasyManua.ls Logo

nal von minden Reactif Touch - Page 63

nal von minden Reactif Touch
88 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
59
Cau on: Any solu ons other than those listed above may damage the test table
and insert.
2. Place the insert and/or test table into the solu on, making sure the white calibra -
on bar on the test table remains above the liquid level.
Cau on: Be sure the solu on does not come in contact with the white calibra on bar.
Do not cover the container while the test table is soaking. .
3. Soak the table and insert for a minimum of 2 minutes and maximum of 10 minutes.
Do not soak longer than 10 minutes.
4. Rinse the test table and insert thoroughly with water.
5. Dry with a so cloth and replace test table and table insert in the analyzer (as de-
scribed on the previous page).
Cleaning the White Calibra on Bar
For the Reac f Touch Analyzer to perform as intended and provide reliable test results,
it is recommended to periodically check the white calibra on bar on the test table. For
proper use, the white calibra on bar should not be dirty or discolored.
1. Remove the insert from the test table.
2. Remove the test table by pulling it slowly out of the analyzer.
3. Check the white calibra on bar on the test table for dirt or discolora on.
White Calibration Bar