Cau on: Care should be taken not to scratch the white calibra on bar. Instruc ons for
cleaning the white calibra on bar are given later in this sec on.
4. Reinsert the test table into the analyzer by holding the table at the end opposite the
white calibra on bar, with the white calibra on bar facing upwards. Push the test tab-
le fi rmly, but slowly just over halfway into the analyzer. Replace the test table insert.
Disinfec ng the Test Table and Insert
1. Prepare one of the following solu ons in a tall, narrow container to a depth of
about 4 inches (10 cm):
• Precept, Cidex, Theracide, and Amphyl solu ons – prepare according to pro
duct direc ons.
• Household Bleach (5% sodium hypochlorite) – this can be used either full
strength or diluted with water to as much as 1:20 (i.e., mix 5 mL bleach with
95 mL water for a total of 100 mL).
• Isopropyl Alcohol (70% to 85%) - this can be used full strength.
Cau on: Rinse away all bleach residues, as remaining bleach will aff ect several of the
reagent pad chemistries.