Cau on: Care should be taken not to scratch the white calibra on bar. Instruc ons for
cleaning the white calibra on bar are given later in this sec on.
4. Reinsert the test table into the analyzer by holding the table at the end opposite the
white calibra on bar, with the white calibra on bar facing upwards. Push the test tab-
le fi rmly, but slowly just over halfway into the analyzer. Replace the test table insert.
Disinfec ng the Test Table and Insert
1. Prepare one of the following solu ons in a tall, narrow container to a depth of
about 4 inches (10 cm):
Precept, Cidex, Theracide, and Amphyl solu ons – prepare according to pro
duct direc ons.
Household Bleach (5% sodium hypochlorite) – this can be used either full
strength or diluted with water to as much as 1:20 (i.e., mix 5 mL bleach with
95 mL water for a total of 100 mL).
Isopropyl Alcohol (70% to 85%) - this can be used full strength.
Cau on: Rinse away all bleach residues, as remaining bleach will aff ect several of the
reagent pad chemistries.