54
nal von minden GmbH · Manual
Reac f Touch
Note: The test number ranges from 200001 to 299999.
Press to exit without running a QC Test.
Press
a er making sure the strip being used is the same as that shown on the
screen. A test screen will be shown as seen below.
Cau on: Use only Reac f Urinalysis Reagent Strips for proper func oning and accurate
results. If it is the fi rst  me running the QC Test, please refer to Sec on 7
Analyzer Setup: Test Se ngs.
Press to cancel the test. If canceled, when the QC Lockout is enabled, the analyzer
will change the test mode to STAT. An “E” will be displayed a er the test number on all
printouts to show QC Tests were out of date and pending. If the QC Lockout is disabled,
the analyzer will have no response.