55
The analyzer will run a Level 1 QC Test fi rst. Refer to Sec on 6 Specimen Test for details
on tes ng. A er tes ng, the analyzer will display the tes ng results.
If any parameter is out of range, it will be highlighted and marked with an *.
QC Test Result
Type of Strip 10U
Date/Time 2012-1-2 8:00:00
Test Number 200001
Level 1 Test Pass
LEU - Neg.
*NIT - Neg.
URO - 0,2 mg/dl
PRO - Neg.
pH 6.0
BLUT - Neg.
SG 1.015
KET - Neg.
BIL - Neg.
GLU - Neg.
QC Test Result
Type of Strip 10U
Date/Time 2012-1-2 8:02:00
Testnummer 200002
Level 2 Test Pass
LEU 2+ 125 Leu/µl
*NIT + Pos.
URO 2+ 4 mg/dl
PRO 3+ 300 mg/dl
pH 8.0
BLUT 3+ 200 Ery/µl
SG 1.025
KET 1+ 15 mg/dl
BIL 2+ 2 mg/dl
GLU 2+ 500 mg/dl