No. Descrip on Ac on
1
Low ba ery power Replace the ba eries. See Sec on 10, Changing Ba eries.
2
Failure of test table
1. Make sure that the test table is in place. Move the test table in or out
of the instrument slightly to reposi on the test table.
2. If the error remains, with the instrument powered on, unplug the pow-
er cord from rear of instrument and plug it back in. Turn the instrument
on by pressing the gray power bu on.
3. If the error remains with the test table in place, contact your local
Customer Service and Support Center.
3 No printer paper
Replace the printer paper. See Sec on 3, Loading the Printer Paper or Label
Roll.
4
Failure of automa c
calibra on
Clean the calibra on strip. If the error remains a er cleaning, contact your
local Customer Service and Support Center.
5 Incorrect strip type
Ensure that the strip type selected in the Instrument Setup matches with
the strip being used. If the correct type of strip is being used, check the
instrument opera on by running another test using: a) a yellow and clear
specimen, or b) Quality control.
6 Missing strip Repeat the test ensuring that the strip is posi oned on the test table.
7 Misplaced strip
Repeat the test ensuring that the strip is correctly posi oned on the test
table. If the error remains, examine the test table insert to ensure the small,
white line located near the p of the strip (on strip side of insert) is present
and not damaged. If this line is damaged or missing, contact your local Custo-
mer Service and Support Center
8
Inverted strip posi -
oned on the test table
Repeat the test ensuring that the strip is correctly posi oned on the test
table.
9 Tilted strip
Repeat the test ensuring that the strip is correctly posi oned on the test
table.
10 Dry strip Repeat the test ensuring that the strip has been immersed in specimen.
11
Failure to fi nd end
of strip
Repeat the test ensuring that the strip is correctly posi oned on the test
table.
Warranty
The Reac f Touch Urine Analyzer has a one-year warranty period. This warranty is
designed to reduce the cost associated with repairing systems that exhibit malfunc ons
due to defects in materials and/or workmanship during the warranty period. One year
warranty starts from the date of original purchaser’s receipt of the product from nal von
minden GmbH. Use the Warranty Registra on Card provided with the instrument to re-
gister the warranty. To obtain assistance during the warranty period, please contact your
local technical support provider or distributor.
List of Errors and Advisory Messages