14-8 Remote control
?FSF n This query command #?FSF n * sends back the complete record in the format as
described in separate document tagged as n. If n =0 then the record is the one just
measured.
In case no records are available the reply is:
FSA= N/A
Otherwise the 9010 sends a Little Endian 32 bit long representing the size of
the array the will follow
?IDN This query command #?IDN* sends back a string containing information about
model, release and date of firmware.
Note that two <LF> are appended to the string before the terminator
Example of reply: IDN=9010-FW - 1.12 20/10/059<LF><LF>
?MAF This query command #?MAF* sends back a string expressing the tuning
frequency, in exponential notation, for manual mode. Unit is fixed MHz.
Example of reply: MAF= 1.500000e+07 which means that the tuned frequency in
manual mode is 15MHz
?MAT This query command #?MAT* sends back a string expressing the Attenuator
status for Manual mode. The reply is made of 2 fields:
• MAT=AUTO (9010 automatically selects the most suitable attenuator to
get best dynamic range) or MAT=MAN ( the operator selects manually the
attenuator)
• Value (in dB) of current attenuator
Example of reply: MAT=AUTO; 20 which means that the Attenuator is now
switched to 20dB automatically by 9010
?MHT This query command #?MHT* sends back a string expressing the HOLD TIME for
manual mode. Unit is fixed ms.
Example of reply: MHT= 1000 ms which means that the hold time in manual mode
is 1000ms (or 1s)
?MIL This query command #?MIL* sends back a string telling whether MIL Filters are
available.
Reply is:
• MIL=N/A if MIL Filters are NOT available
• MIL=OK if MIL Filters are available
?MPA This query command #?MPA* sends back a string expressing the Preamplifier
status for Manual mode. The reply is either:
• MPA=Off (Preamplifier in Manual Mode is OFF)
• MPA =On (Preamplifier in Manual Mode is ON)
?MPS This query command #?MPS* sends back a string expressing the Preselector
status for Manual mode. The reply is either:
• MPS=Off (Preselector in Manual Mode is OFF)
• MPS =On (Preselector in Manual Mode is ON)
?PLM This query command #?PLM* sends back a string expressing the Pulse Limiter
status. The reply is either:
• PLM =Off (Pulse Limiter is OFF)
• PLM =On (Pulse Limiter is ON)