14-24 Remote control
SPLM a
This setting command switches the Pulse Limiter; according the parameter (a) which
can be either On or OFF.
The reply is PLM =OK which acknowledges the command has been granted or PLM
=SERR if the command has been ignored.
Example: # PLM ON*
Example: # PLM OFF*
SRBW b
This setting command sets RBW. The argument (b) should be a string representing
the RBW as follows:
o 0 RBW is Automatic so that the 9010 automatically selects the
suitable RBW according to the tuned frequency
o 1 300kHz
o 2 100kHz
o 3 30kHz
o 4 10kHz
o 5 3kHz
o Only for 9010
6 9kHz (Cispr 16-1)
7 200Hz (Cispr 16-1)
11 10kHz (MIL-STD-461) (only when available)
22 1kHz (MIL-STD-461) (only when available)
23 100Hz (MIL-STD-461) (only when available)
24 10Hz (MIL-STD-461) (only when available)
o Only for Downconverter
9 1MHz (Cispr 16-1)
10 120kHz (Cispr 16-1)
12 9kHz (Cispr 16-1) ; No QPeak available
The reply is RBW =OK which acknowledges the command has been granted or
RBW =SERR if the command has been ignored.
Example: #S RBW 3* sets the RBW to 30kHz and disables Auto RBW mode
Example: #S RBW 0* turn the RBW mode in Automatic
SSRT f
This setting command sets the Start frequency for Sweep mode. The string (f) can
be in exponential form and should be expressed in Hz. The reply is SRT=OK which
acknowledges the command has been granted or SRT =SERR if the command has
been ignored.
Example: #SSRT 150e3* Sets the Start frequency, for Sweep mode, to 150kHz
SSTP
This setting command stops abruptly a running sweep in Analyzer mode. The reply
is always STP=OK, which acknowledges the command has been granted
STAT b
This setting command sets MinAtt to the value indicated by (b) which range from
0dB to the maximum the 9010/9030/9060/9180 can set and must be in step of 5dB.
The reply is TAT=OK which acknowledges the command has been granted or TAT
=SERR if the command has been ignored.
Note: In Sweep mode the Attenuator is always forced to AUTO
Example: #STAT 10*
STGF f
This setting command sets the tuning frequency for 9010 Built-In Generator when is
not tracking (STGT OFF). The string (f) can be in exponential form and should be
expressed in Hz.
Frequency should range between 1 kHz and 50MHz inclusive.
The reply is TGF=OK which acknowledges the command has been granted or TGF
=SERR if the command has been ignored.
Example: #STGF 150e3* Sets the tuning frequency, for 9010 Built-In Generator, to
150kHz