Chapter 4 Connecting and Using Accessories and Transducers
Getting Started with SCXI 4-22 www.ni.com
Excitation Channel Signal Connections on the
SCXI-1122
The SCXI-1122 has a voltage (VEX) and a current excitation (IEX)
channel, which are available at the front connector. In addition, VEX/2 is
available for half-bridge and quarter-bridge transducers. Both channels are
isolated from earth ground up to 450 V
rms
working common-mode voltage.
Notice that the voltage and current excitation channels are electrically
isolated from each other but do not provide a safety isolation between them.
Warning
Exceeding the overvoltage protection or isolation rating on the excitation output
can damage the SCXI-1122, the SCXIbus, and the DAQ device. National Instruments is
not liable for any damages or injuries resulting from such signal connections.
Excitation Level
Each excitation channel of the SCXI-1122 has one level:
Current excitation1mA
Voltage excitation3.333 V
It is important that you do not exceed the maximum permissible load of
each channel as listed in Table 4-4.
Table 4-4. SCXI-1122 Maximum Load per Excitation Channel
Excitation Level Maximum Load
3.333 V
1mA
225 mA
5k

Table of Contents

Other manuals for National Instruments SCXI Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the National Instruments SCXI Series and is the answer not in the manual?

National Instruments SCXI Series Specifications

General IconGeneral
BrandNational Instruments
ModelSCXI Series
CategoryControl Unit
LanguageEnglish

Related product manuals