EasyManua.ls Logo

natus Echo-Screen III Pro - Page 50

natus Echo-Screen III Pro
54 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
026057 RevE 46 © 2020 Natus Medical Incorporated
Appendix A: Electromagnetic Compatibility (EMC) Echo-Screen III Pro Instructions for Use
Table 8 – Immunity Test Levels - Signal Input / Output Parts Port
Table 9 - Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless
communications equipment
Phenomenon
Basic EMC
Standard
Immunity Test Levels - Professional Healthcare
Facility Environment
Electrostatic Discharge IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
Electrical Fast
Transients/Bursts
IEC 61000-4-4
± 1 kV
100 kHz repetition frequency
Surges Line-to-Ground
(Common Mode)
IEC 61000-4-5 ± 2 kV
Conducted Disturbances
Induced by RF Fields
IEC 61000-4-6
3 V
0,15 MHz – 80 MHz
6 V in ISM bands between 0,15 MHz and 80 MHz
80% AM at 1 kHz
Test
Frequency
(MHz)
Band
(MHz)
Service Modulation
Maximum
Power (W)
Distance
(m)
Immunity
Test
Level
(V/m)
385
380 –
390
TETRA 400
Pulse modulation
18 Hz
1,8 0,3 27
450
430 –
470
GMRS 460,
FRS 460
FM
± 5 kHz deviation
1 kHz sine
20,328
710
704 –
787
LTE Band
13, 17
Pulse modulation
217 Hz
0,2 0,3 9745
780
810
800 –
960
GSM
800/900
TETRA 800
iDEN 820
CDMA 850
LTE Band 5
Pulse modulation
18 Hz
20,328
870
930
1,720
1,700 –
1,990
GSM 1800
CDMA 1900
GSM 1900
DECT
LTE Band 1,
3, 4, 25;
UMTS
Pulse modulation
217 Hz
20,328
1,845
1,970
2,450
2,400 –
2,570
990Bluetoot
h
WLAN
802.11 b/g/n
RFID 2450
LTE Band 7
Pulse modulation
217 Hz
20,328

Related product manuals