© 2020 Natus Medical Incorporated 47 026057 RevE
Appendix A: Electromagnetic Compatibility (EMC) Echo-Screen III Pro Instructions for Use
NOTICE: Deviations from the collateral standard were performed per IEC 60601-2-40
NOTICE: All necessary instruction for maintaining compliance with regard to EMC can be found
in the general maintenance section in this instruction. No further steps are required.
To ensure compliance with the EMC requirements as specified in IEC 60601-1-2, it is
essential to use only the following accessories:
NOTICE: Conformance to the EMC requirements specified in IEC 60601-1-2 is ensured if the
cable types and cable lengths are as specified above:
5,240
5,100 –
5,800
WLAN
802.11 a/n
Pulse modulation
217 Hz
0,2 0,3 95,500
5,785
ITEM MODEL MANUFACTURER
Cable length
[meters]
Echo-Screen III OAE Probe Cable 012625 Natus Medical 1.2 M
Echo-Screen III ATA Cable 010999 Natus Medical 1.2 M
Echo-Screen III PCA Cable 009149 Natus Medical 1.2 M
Universal AC Adapter 013401 GlobTek 1.83 M
USB Cable, A to Micro B 011029 L-Com 2 M
Test
Frequency
(MHz)
Band
(MHz)
Service Modulation
Maximum
Power (W)
Distance
(m)
Immunity
Test
Level
(V/m)