Calibrate the Scale 
1.  Press the Scale Zero Button to display the Cal and PAr values. Record these values. 
2.  Install the Produce Guard if one is included then install the Top Plate. Raise the front edge 
of the Top Plate enough to access the Calibration Switch and press it, then quickly lower the 
Top Plate. The display should show Ready C-00 kg (00 lb.). 
3.  Follow the sequence of steps in the following chart. 
 
Display  Add Weight  Remove Weight 
Ready C-2.5 kg (05 lb.)  2.50 kg (5.00 lb.)   
Ready C-05 kg (15 lb.)  2.50 kg (10.00 lb.)   
Ready C-10 kg (30 lb.)  5.00 kg (15.00 lb.)   
Ready C-00 kg (00 lb.)    10.00 kg (30.00 lb.) 
Ready 0.000 kg (0.00 lb.)     
 
Verify the Calibration 
The scale accuracy test meets government requirements of testing the accuracy of the scale after 
performing a scale calibration. It contains a series of four tests that must be run in the continuous 
sequence given. 
Increasing Load Test 
This test checks the scale’s accuracy when incrementally adding weight to the center of the top 
plate. Use weights that correspond to the RealScan 7875 weight feature. 
 
Step  Weight Feature  Add Weight  Remove Weight  Display Result 
1  9.995 kg  0.1 kg    0.1 ± 0.00 kg 
  13.995 kg  0.1 kg    0.1 ± 0.00 kg 
 
30.0 lb.  0.2 lb.    0.2 ± 0.00 lb. 
2 
9.995 kg  2.5 kg  0.1 kg  2.5 ± 0.00 kg 
  13.995 kg  2.5 kg  0.1 kg  2.5 ± 0.00 kg 
 
30.0 lb.  5.0 lb.  0.2 lb.  5.0 ± 0.00 lb. 
3 
9.995 kg  2.5 kg    5.0 ± 0.005 kg 
  13.995 kg  4.5 kg    7.0 ± 0.005 kg 
 
30.0 lb.  5.0 lb.    10.0 ± 0.01 lb. 
4 
9.995 kg  2.5 kg    7.5 ± 0.005 kg 
  13.995 kg  3.0 kg    10.0 ± 0.005 kg 
 
30.0 lb.  10.0 lb.    20.0 ± 0.01 lb. 
5 
9.995 kg  2.495 kg    9.995 ± 0.005 kg 
  13.995 kg  3.995 kg    13.995 ± 0.005 kg 
 
30.0 lb.  10.0 lb.    30.0 ± 0.01 lb. 
 
Note: Do NOT remove any weight from the Top Plate.