NCR RealScan 7875-7000/8000 Installation Guide 
  06/03  AIP-00342    Release B 
22 of 56 
Step 5 – Calibrate the Scale 
Note: You MUST calibrate the scale when you install a RealScan 7875-80xx. To be in 
compliance with governmental weights and measures regulations, you must be certified to 
calibrate the scale. 
Note: Calibration information is NOT sent to the host terminal. If your RealScan 7875 does not 
have voice enabled or does not have a remote display, you must use a Field Service Calibration 
Display. 
Note: Be sure the Interface Cable between the RealScan 7875 and the host terminal is 
disconnected during scale calibration. Some host terminals can cause interference that can 
invalidate the scale calibration. 
You can calibrate the scale after power has been supplied for 30 minutes if the ambient 
temperature has been 20° C (68° F) for at least 24 hours. If this condition is not met, then the 
scale must be on for at least 6 hours before you can calibrate it. Also, you must use a certified 
weight set to calibrate and certify the scale. 
The RealScan 7875 maintains an audit trail of scale calibration and programming activity. The 
audit trail records the number of times the scale has been calibrated. You can display the audit 
trail count on a Remote 7825 or Field Service Calibration Display by pressing the Scale Zero 
Button on the Operator Display Panel. The display alternates between Cal xxx and PAr xxx. 
Set Scale AC Voltage Frequency 
The RealScan 7875-8000 units contain a scale that must be set for the correct AC voltage 
frequency before the scale is calibrated. 
1.  Turn on the AC Power by turning on the circuit breaker. 
2.   Set the AC voltage frequency by scanning the following sequence of programming tags. 
These must be the first tags scanned after applying power to the unit. 
 
60 Hz AC Voltage 
Programming Mode, Hex 3, Hex 0, Hex F – Default setting 
50 Hz AC Voltage 
Programming Mode, Hex 3, Hex 0, Hex E 
Exercise the Scale 
You must exercise the scale before performing a calibration. Add and remove weight to your 
scale in the order given in the following chart. This is a building procedure in which you start 
with no weight (0) and sequentially add and remove weight to obtain the values shown in the 
chart. Go through this procedure four times. 
 
Weight 
Feature 
 
Total Weight on Top Plate 
9.995 kg  0.0 kg  2.5 kg  5.0 kg  10. 0 kg  5.0 kg  2.5 kg  0.0 kg 
13.995 kg  0.0 kg  2.5 kg  5.0 kg  15.0 kg  5.0 kg  2.5 kg  0.0 kg 
30.0 lb.  0.0 lb.  5.0 lb.  15.0 lb.  30.0 lb.  15.0 lb.  5.0 lb.  0.0 lb.