Model H1301 Recovery and Recycling System
Installation, Operation and Maintenance Manual
Revision-L September, 2019
Page 89 of 105
VII. System Maintenance Procedures
P. Cooling System Controller Setting (continued)
3. Cooling System Temperature Controller Setup Parameters:
The cooling system temperature controller setup parameters have been factory
programmed and do not need to be adjusted with the exception of the offset calibration
OFFS. The below listing is for reference only or may need to be completed if installing a
new controller. Set up parameters not listed below are not applicable for the machine
operation and may be set to any available selection. Changing any of the Configuration
Menu parameters will automatically set all of the Set Up parameters to the default
settings. A detailed description of the controller setup parameter function may be found in
the enclosed temperature chamber manual.
To access the Configuration Mode:
a. Press and hold the Function-key then press and hold with the UP arrow key within
30 seconds after power on: the controller will enter the Select Mode. Press the
Down Arrow-key until the upper display indicates ConF. Press the Function-key:
the lower line should display ULOC.
b. Use the UP arrow key to enter the number 20 into the upper display line then press
the Function-key. The controller should now enter the configuration programming
mode.
c. Use the Function-key to step through the menu items.
Configuration Mode Factory Settings:
Prompt Setting
inPE tF
rul 80
rLL -80
CtYP SnGL
CtrL dir
ALA1 nonE
ALA2 nonE
LAEn diSA
Inhi none
USE 1 Pri
diSP 1
CLoc 20