EasyManuals Logo

Nikon ECLIPSE Ni-E User Manual

Nikon ECLIPSE Ni-E
186 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #50 background imageLoading...
Page #50 background image
Chapter 2 Microscopy Operations
36
Epi-fluorescence
Microscopy
Preparation Focusing Microscopy
Epi-fluorescence
Microscopy Procedure
12345 67891011121314
5
Turn on the mercury lamp.
See your illuminator's manual for details.
INTENSILIGHT
C
-
HGFIE
LAMP
POWER
RUN TIME hrs.
Mercury lamp ON
6
To observe under a different magnification
Press the OBJ CW/CCW button to bring the objective
with the desired magnification into the optical path.
Pressing the OBJ CW button turns the nosepiece by one
address in the clockwise direction (as viewed from
above), while the OBJ CCW button turns the nosepiece
in the counterclockwise direction.
Oil immersion
When using an oil immersion type objective, apply Nikon
specified non-fluorescence immersion oil between the
specimen and the objective. (See Chapter 3 “14
Oil/Water Immersion”.)
Motorized nosepiece toggling
Two objectives can be alternated by specifying an
objective pair. Objectives can be switched using the
[OBJ CW/CCW] button on the microscope. (See
Chapter 3 “11.1 Motorized Nosepiece Operation”.)
Bringing desired objective into optical path
PF_ 40x____Z:_____0.000um
UV-2B
'
C ____ND_ 1
'
C __EPI:C
[NOSEPIECE] button, sub screen: Desired objective
Power indicator (green)
ON
Power switch ON

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Nikon ECLIPSE Ni-E and is the answer not in the manual?

Nikon ECLIPSE Ni-E Specifications

General IconGeneral
BrandNikon
ModelECLIPSE Ni-E
CategoryMicroscope
LanguageEnglish

Related product manuals