Chapter 2 Microscopy Operations
60
Differential
Interference Contrast
Microscopy
Preparation
Focus and Optical
System Adjustment
Microscopy
Differential Interference
Contrast Microscopy
12345678910111213141516 17 18 19 20 21 22 23
17
Bring the DIC module (on the condenser) into the optical path.
Press the FUNCTION button 1/2 to turn the condenser
turret and bring the DIC module appropriate for the
objective into the optical path.
Button 1 is for counterclockwise rotation and button 2 is
for clockwise rotation.
Display DIC module [N1], etc. on the NI-E's display
panel. The turret's DIC module enters the optical path in
this state.
Choose the DIC module appropriate for the intended
microscopy purpose and the objective used. (See
Chapter 3 “17 Differential Interference Contrast
Microscopy” - “17.2 Using Optical Elements”.)
IN
ND
8
ND
32
NCB
11
OUT
Bring DIC module for the condenser into optical path
PF_ 40x____Z:__1234.567um
N1____AS30.0mm__FS30.0mm
Example
[CONDEN.] button, sub screen: [N1] (Example)
Optical module
in optical path
[N1]