Chapter 2 Microscopy Operations
61
Differential
Interference Contrast
Microscop
Preparation
Focus and Optical
System Adjustment
Microscopy
Differential Interference
Contrast Microscopy
12345678910111213141516 17 18 19 20 21 22 23 ■ ■
18
Adjust the aperture diaphragm
Press the DIA aperture diaphragm button to adjust the
aperture. The aperture should be adjusted to about 70 to
80% of the numerical aperture of the objective. (See
Chapter 3 “8 Adjusting the Aperture Diaphragm”.)
Appropriate size of the aperture
diaphragm
Normally, the appropriate size of the aperture
diaphragm is 70 to 80% of the numerical aperture of
the objective. Since an excessively small aperture
diaphragm opening will degrade the image
resolution, Nikon does not recommend setting the
aperture diaphragm to less than 60% of the
numerical aperture of the objective.
Adjustment timing for the aperture
diaphragm
Be sure to adjust the aperture diaphragm each time
you change the objective.
ND
8
ND
32
OUT
IN
NCB
11
Adjusting aperture diaphragm
PA_ 40x____Z:__1234.567um
N1___AS12.8mm___FS30.0mm
Appropriate size of the aperture diaphragm
[A. STOP] button, sub screen: DIA aperture
diaphragm adjustment
19
Focus on the sample
Focus on the sample by rotating the focus knob. (See
Chapter 3 “4.1 Proper Focusing Procedure”.)
Focusing using the focus knob
PA_ 40x____Z:__1235.555um
N1___AS12.8mm___FS30.0mm
Z-axis coordinate
perture
dia
hra
m ima
e
Pupil of objective
70~80
100