Leaf User Manual
P/N: 97-4000, Revision 08, Release date: 2020-11-10 Page 34 of 85
Intraoperative Instructions
Slit Lamp Treatment Procedure
Perform the following procedure in conjunction with the instructions in the Slit Lamp
operator manual.
1.
Connect the SLA to the slit lamp, as described in the “Connecting the Slit
Lamp Adapter (SLA)” section.
2.
Verify that the slit lamp eyepieces are adjusted to your settings, as described in
the slit lamp operator manual.
3.
Position the patient at the slit lamp with their chin on the chin rest and
forehead pressed firmly against the head rest.
4.
Select the laser treatment spot diameter, exposure time and beginning
treatment power level.
5.
Position the contact lens on the patient's eye.
6.
Select READY mode. The aiming beam will turn on.
7.
Adjust the aiming beam intensity.
8.
Focus the slit lamp and observe the red aiming beam imaged on the patient's
eye. Verify that the laser spot is round and undistorted. Establish proper
placement of the laser beam with the slit lamp joystick or micromanipulator.
9.
Depress and hold down the footswitch to deliver the treatment laser beam to the
tissue
10.
Adjust the laser treatment power as needed for therapeutic effect.
11.
Prior to continuing treatment, verify that the power and other parameters are
within acceptable ranges.
12.
When treatment is complete, press “END” to return the main screen to default
parameters.
Please Note: Treatment may be interrupted at any point by releasing the footswitch.
NOTE
Always place the system in STANDBY mode if there is a prolonged pause in treatment.
If the system is in READY mode and remains idle for 5 minutes, it automatically
reverts to STANDBY mode. To resume treatment, place the system back into READY
mode.
If the system remains idle for 30 minutes, it will END the current treatment and return
all parameters to the default position. The pulse count will retain the same pulse count
of the prior treatment and can only be reset by pressing the Reset button on the
user interface.