P/N: 97-4000, Revision 08, Release date: 2020-11-10 Page 47 of 85
Code
Message
Critical
Clearable
Warning
Description
Action
1-42
Interlock Door Open
X
Door Interlock
connection is OPEN
Close Room Door
Check Door interlock
Connection
Turn Key OFF, then ON
1-43
Interlock Key Switch
X
Key Switch Contacts
Failure
Turn Key OFF, then ON
Call service if error persists.
1-44
Interlock Relay
Failure
X
Key Switch Relay
Contacts Failure
Turn Key OFF, then ON
Call service if error persists.
1-45
Emergency Shut
Down
X
Emergency Off
Switch Engaged
Turn Key OFF, then ON
Call service if error persists.
1-61
to
1-91
Analog Signal
Integrity
X
Hardware error
Turn Key OFF, then ON
Call service if error persists.
2-1
Failed to create
Bluetooth adapter
X
Bonding failure
Turn Key OFF, then ON &
reboot tablet
Call service if error persists.
2-2
Failed to create
Bluetooth LE
Scanner
X
Bonding failure
Turn Key OFF, then ON &
reboot tablet
Call service if error persists.
2-3
Failed to create
Bluetooth GATT
X
Bonding failure
Turn Key OFF, then ON &
reboot tablet
Call service if error persists.
2-4
Failed to initiate
Create Bond
X
Bonding failure
Turn Key OFF, then ON &
reboot tablet
Call service if error persists.
2-5
Failed to initiate Set
Pin
X
Bonding failure
Turn Key OFF, then ON &
reboot tablet
Call service if error persists.
2-6
Failed to create
Bluetooth device
X
Bonding failure
Turn Key OFF, then ON &
reboot tablet
Call service if error persists.
2-7
Failed to initiate
discover services
X
Bonding failure
Turn Key OFF, then ON &
reboot tablet
Call service if error persists.
2-8
Failed to discover
services
X
Bonding failure
Turn Key OFF, then ON &
reboot tablet
Call service if error persists.
2-9
Failed to create
Bluetooth GATT
Services
X
Bonding failure
Turn Key OFF, then ON &
reboot tablet
Call service if error persists.
2-10
Failed to bond to
correct device
X
Bonding failure
Turn Key OFF, then ON &
reboot tablet
Call service if error persists.
2-11
Failed to validate
Services
X
Bonding failure
Turn Key OFF, then ON &
reboot tablet
Call service if error persists.
2-12
Characteristic
indication
unexpected
X
Bonding failure
Turn Key OFF, then ON &
reboot tablet
Call service if error persists.
2-13
Characteristic read
empty queue
X
Bonding failure
Turn Key OFF, then ON &
reboot tablet
Call service if error persists.