P/N: 97-4000, Revision 08, Release date: 2020-11-10 Page 49 of 85
Code
Message
Critical
Clearable
Warning
Description
Action
2-32
Failed Descriptor
Write Async
X
Verify hardware is powered
on and within range
Call service if error persists.
2-33
Failed characteristic
read sync
X
Verify hardware is powered
on and within range
Call service if error persists.
2-34
Failed characteristic
read Async
X
Verify hardware is powered
on and within range
Call service if error persists.
2-35
Failed characteristic
write sync
X
Verify hardware is powered
on and within range
Call service if error persists.
2-36
Failed characteristic
write Async
X
Verify hardware is powered
on and within range
Call service if error persists.
2-37
Action timed out
X
Verify hardware is powered
on and within range
Call service if error persists.
3-1
Failed to create USB
manager
X
Cable
communication error
Turn Key OFF, then ON &
reboot tablet
Call service if error persists.
3-2
Read when closed
X
Cable
communication error
Turn Key OFF, then ON &
reboot tablet
Call service if error persists.
3-3
Write failed
X
Cable
communication error
Turn Key OFF, then ON &
reboot tablet
Call service if error persists.
3-4
Read failed
X
Cable
communication error
Turn Key OFF, then ON &
reboot tablet
Call service if error persists.
3-5
Reading timed out
X
Cable
communication error
Turn Key OFF, then ON &
reboot tablet
Call service if error persists.
3-6
Bad SOT
X
Cable
communication error
Turn Key OFF, then ON &
reboot tablet
Call service if error persists.
3-7
Bad escape
X
Cable
communication error
Turn Key OFF, then ON &
reboot tablet
Call service if error persists.
3-8
Received NACK
X
Cable
communication error
Turn Key OFF, then ON &
reboot tablet
Call service if error persists.
3-9
Received Busy
X
Cable
communication error
Turn Key OFF, then ON &
reboot tablet
Call service if error persists.
3-10
Received ACK
X
Cable
communication error
Turn Key OFF, then ON &
reboot tablet
Call service if error persists.
3-11
Received READ
X
Cable
communication error
Turn Key OFF, then ON &
reboot tablet
Call service if error persists.