Leaf User Manual
P/N: 97-4000, Revision 08, Release date: 2020-11-10 Page 52 of 85
Failed to get
response over serial
Verify bonding cable is
connected
Turn Key OFF, then ON &
reboot tablet
Call service if error persists.
Verify bonding cable is
connected
Turn Key OFF, then ON &
reboot tablet
Call service if error persists.
Verify laser is turned ON
Turn Key OFF, then ON &
reboot tablet
Call service if error persists.
System mode not
ready in time
Turn Key OFF, then ON
Call service if error persists.
Spot Size Not Ready
In Time
Turn Key OFF, then ON
Call service if error persists.
Pressed footswitch
while in Standby
Place unit into Treat mode
Call service if error persists.
Pressed footswitch
while entering Treat
Wait until after 3 second
delay entering treat
Call service if error persists.
Pressed footswitch
while laser is
disabled
Wait after modifying
parameters before pressing
footswitch
Call service if error persists.
Turn Key OFF, then ON
Call service if error persists.
Smaller spot size has resulted
in increased energy density.
Verify energy density and
power before firing laser.
Selected power factor
resulted in a power higher
than the maximum. Power
was set to the maximum
allowed.