P/N: 97-4000, Revision 08, Release date: 2020-11-10 Page 52 of 85
Code
Message
Critical
Clearable
Warning
Description
Action
7-2
Failed to get
response over serial
X
Laser Error
Verify bonding cable is
connected
Turn Key OFF, then ON &
reboot tablet
Call service if error persists.
7-3
Failed to bond
X
Laser Error
Verify bonding cable is
connected
Turn Key OFF, then ON &
reboot tablet
Call service if error persists.
7-4
Failed to connect
X
Laser Error
Verify laser is turned ON
Turn Key OFF, then ON &
reboot tablet
Call service if error persists.
7-5
System mode not
ready in time
X
Laser Error
Turn Key OFF, then ON
Call service if error persists.
7-6
Spot Size Not Ready
In Time
X
Laser Error
Turn Key OFF, then ON
Call service if error persists.
7-7
Pressed footswitch
while in Standby
X
Laser Error
Place unit into Treat mode
Call service if error persists.
7-8
Pressed footswitch
while entering Treat
X
Laser Error
Wait until after 3 second
delay entering treat
Call service if error persists.
7-9
Pressed footswitch
while laser is
disabled
X
Laser Error
Wait after modifying
parameters before pressing
footswitch
Call service if error persists.
7-10
Lost connection
X
Laser Error
Turn Key OFF, then ON
Call service if error persists.
8-1
Reduced Spot Size
X
Laser Error
Smaller spot size has resulted
in increased energy density.
Verify energy density and
power before firing laser.
8-2
Power Factor Railed
X
Laser Error
Selected power factor
resulted in a power higher
than the maximum. Power
was set to the maximum
allowed.