Leaf User Manual
P/N: 97-4000, Revision 08, Release date: 2020-11-10 Page 61 of 85
If the system becomes unresponsive at any time other than during laser emission, do
not press the emergency laser stop button. Instead, turn the key to the OFF position.
Wait at least one minute before restarting the system using the key switch.
If the control panel will not connect to the laser for more than 60 seconds during
system start-up, verify that the power indicator LED on the side of the laser is
illuminated. If it is not illuminated, then check the power cord and cables from the wall
outlet. If the control panel remains blank, turn off the system with the key; verify that all
control panel cables are plugged in and fully seated; and then restart the system. If the
system still has no power, turn off the system and contact service.
It may take longer for the equipment to achieve a ready state in a low-
temperature environment.
It is the responsibility of the physician to verify that the aiming beam spot visualized
through the slit lamp is the expected size. If the aiming beam size appears
inappropriate or distorted, do not proceed with treatment. Readjust the slit lamp
focus. If the problem persists, contact Norlase service or your authorized Distributor.
It is the responsibility of the physician to select the appropriate power and treatment
location. The lowest practical setting should always be used to achieve the desired
clinical outcome.
Do not use a wet cloth to clean the control panel screen. Doing so may damage the
screen.