5-7
5 Maintenance
5. Maintenance
To perform a Long-Term Shutdown and ESM Flow Path Purge:
1.  Remove the ESM Reagent Cartridge from the pack bay, while keeping all other 
ow components installed.
2.  Install  the  6-port  Flush  Fixture  into  the  reagent  pack  bay.  Make  sure  that  the 
needle tments at the rear of the xture engage with the needle shrouds in the 
back of the pack bay.
3.  Locate the length of tubing labeled “W” (Waste) on the xture and place the end 
of this waste tubing into an empty waste collection beaker.
4.  Gather all the remaining lengths of tubing on the xture and place them into a 
beaker containing at least 250 mL of DI H2O. Make sure the ends of each line of
tubing are submerged.
5. Select the ESM Long-Term Shutdown button and then select Start to begin the
Flow Path Flush sequence.
6.  The ESM will drain all residual reagent from its tubing and will ush the owpath 
with the DI water for several minutes.
7.  When the sequence is complete, remove each piece of ush xture tubing from 
the DI water beaker; leave the “W” tubing in the waste collection beaker.
8.  Select Start again to purge the residual DI H2O from the ow path and ush it 
with air.
9.  When the second sequence is complete, remove the ush xture and power down 
the ESM using the rear power switch.
NOTE: The Long-Term Shutdown procedure for the ESM can be run at the same time
as the Long-Term Shutdown procedure for the FLEX2, but the ESM Shutdown
Procedure must be started rst.
5.1.6 Cleaning FLEX2 + ESM Surfaces
Nova  Biomedical  Corporation  recommends  using  70%  Reagent  Alcohol  (V/V)  or 
Isopropyl Alcohol (IPA) for cleaning the various FLEX2 + ESM surfaces or components 
when required. Use a lint-free cloth or Kimwipe® lightly dampened with the cleaning 
reagent to wipe down surfaces. Never spray or pour reagent directly onto or into the 
analyzer or ESM. Once wiped down, all residual uid should be dried with a lint-free 
cloth or Kimwipe®. 
WARNING: Vapor from other cleaning reagents used within the lab may be corrosive to
the BioProle FLEX2 + ESM and could result in damage to the system. Use
caution and protect the system, as needed, when using reagents that produce
toxic vapors.