EasyManua.ls Logo

NOVA BIOMEDICAL BIOPROFILE FLEX2 - Online Sampling; Analysis Preparation; Sampling Sequence

NOVA BIOMEDICAL BIOPROFILE FLEX2
45 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-8
BioProfile FLEX2 ESM Instructions for Use Manual
4.4 Online Sampling
4.4.1 Analysis Preparation
Prior to sampling, the ambr® queries the FLEX2 for ESM status along with the status 
of the analytical modules included in the selected Sample Type. The ESM must be
connected, initialized, and primed. Further, all consumables
must  be  installed  and  valid  (non-expired),  the  selected 
FLEX2 modules must be available and primed, and key 
parameters  must  be  calibrated  and  passing  QC  (if  QC 
Lockout is enabled) for the analysis sequence to begin. 
The analysis time for an online sample is 6-7 minutes from
the ambr® request to results reporting, depending on the 
congured analysis panel.
4.4.2 Sampling Sequence
1. The ambr® computer sets the FLEX2 to a ‘Busy’state, reserving it for analysis. 
The FLEX2 sequence countdown timer icon will appear in the Status Bar.
2. The ambr® robot arm obtains a full sample from a microculture vessel (400, 450, 
or 500µL from ambr® 15; 675µL from ambr® 250)
3. The ambr® robot arm moves above the ESM sample cup and dispenses 30 µL
of sample into the cup for ow test.
4. The ESM syringe pump drains the test sample from the sample cup and draws
it to the ESM air detector.
5. The ambr® robot arm dispenses the remainder of the sample volume into the
sample cup.
6. The ESM syringe pump drains the cup and draws the sample through the sample
(CF) line to the ESM air detector.
7. The air detector is queried for uid to ensure there is sufcient sample volume.
8. If no uid is detected at the air detector, the ESM syringe draws again and repeats 
Step 7.
9. If uid is detected, FLEX2 sample acquisition proceeds.
10. The FLEX2 sample probe moves to the sample port septum and aspirates the
sample.
11. FLEX2 sample analysis continues based on the selected modules congured in 
the Sample Type.
12. Analysis results appear on FLEX2 User Interface and are stored in FLEX2 Historical
Results. They are automatically transmitted to the ambr® database.
13. The ESM ow path is purged, followed by an automated cleaning cycle.
Figure 4.12 Sampling
Sequence

Related product manuals