EasyManua.ls Logo

NOVA BIOMEDICAL BIOPROFILE FLEX2 - Page 31

NOVA BIOMEDICAL BIOPROFILE FLEX2
45 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-7
4 ESM System Operation
4. ESM System Op.
4.3.1.1 Key Parameters
The  Key  Parameters  function
provides a means of suppressing
analysis of online samples when
specic "key" parameters are not 
available due to calibration failure
or  Quality  Control  failure  (if  QC 
Lockout  is  enabled).  The  Key
Parameters function only applies
to samples supplied to the FLEX2
by the External Sampling Module 
(ESM);  it  does  not  apply  to  the 
Manual  or  Load-N-Go  carousel 
sampling modes. An operator
has the option of identifying
which parameters are "key" when 
creating or modifying a Sample Type.
To congure Key Parameters:
1. From the Sample Type menu (Fig. 4.10) select "True" from the drop down in the 
column labeled "Key" for the parameter of interest.
2. Press the green checkmark to save the Sample Type.
If an online sample is scheduled for analysis using a FLEX2 Sample Type that has Key 
Parameters enabled, all Key Parameters must be available for the analysis to proceed. 
If a  Key Parameter  is unavailable  for any  reason (QC  Lockout included),  any online 
sample analyses with that parameter included in the analysis panel will not run. This
feature prevents the loss of sample material when there may be limited volume.
4.3.1.2 ESM Sampling Volume
In the bottom left corner of the
Sample Type menu is a dropdown
labeled "ESM Volume." For FLEX2 
+ ESM systems, this volume is the 
sample acquisition volume from
the microbioreactor system. For
FLEX2's  integrated  with  ESM15,
there is a choice of 3 different
sample volumes: 400µL, 450µLor 
500µL. For FLEX2's integrated with 
ESM250, there is only one default 
sampling volume: 675 µL. 
NOTE: To determine which volume is optimal for sampling from the ambr® 15, Nova
Biomedical recommends performing a correlation study to test similarity of
results between manual sampling and ESM15 sampling at the three different
acquisition volumes.
Figure 4.10 Congure Key Parameters
Figure 4.11 ESM Sampling Volume

Related product manuals