1.3 System Requirements
Electrical Requirements
Two grounded, 3-wire  receptacles within 5 feet of  the system are required for 
operation.
It is recommended that the BioProle FLEX2 Analyzer and the ESM are connected 
to a 1100 – 1200 VAUniversal Power Supply (UPS) to maintain the power load 
in the event of a power outage.
FLEX2 Fuse Requirements
Operating Voltage: 90 - 264 VAC
Operating Frequency: 50 - 60 Hz
Nominal Power Consumption (without Osmo): 480 Watts
Nominal Power Consumption (with Osmo): 590 Watts
Fuses (2): Slo-Blo (SB) 8A or T8A/250V
ESM
Operating Voltage:  100 to 120, 220 to 240 VAC
Operating Frequency:  50 - 60 Hz
Fuse Rating:  Slo-Blo (SB) 2Aor T8A/250V
Environmental Specications
The BioProle FLEX2 ESM should be operated indoors. Keep the working area 
around the system free of dirt & debris, corrosive fumes, and excessive temperature 
changes.
Ambient operating temperature between 15°C to 30°C (59°F to 86°F)
Altitude up to 2500 meters (8202 Feet)
Relative Humidity of 20 – 85% (without condensation)
Installation Category (II)
Pollution Degree (2)
Dimensions (ESM unit only)
Height:  6.9 in. (17.53 cm) 
Width:  9.7 in. (24.64 cm) 
Depth:  18.2 in. (46.23 cm)
Weight:
8.6 lbs. (3.9 kg) Without Reagent Cartridge
17.8 lbs. (8.1 kg) With Reagent Cartridge