1 Introduction
1. Introduction
1.3 System Requirements
Electrical Requirements
• Two grounded, 3-wire  receptacles within 5 feet of  the system are required for 
operation.
• It is recommended that the BioProle FLEX2 Analyzer and the ESM are connected 
to a 1100 – 1200 VA Universal Power Supply (UPS) to maintain the power load 
in the event of a power outage.
FLEX2 Fuse Requirements
• Operating Voltage: 90 - 264 VAC
• Operating Frequency: 50 - 60 Hz
• Nominal Power Consumption (without Osmo): 480 Watts
• Nominal Power Consumption (with Osmo): 590 Watts
• Fuses (2): Slo-Blo (SB) 8A or T8A/250V
ESM
• Operating Voltage:          100 to 120, 220 to 240 VAC
• Operating Frequency:          50 - 60 Hz
• Fuse Rating:            Slo-Blo (SB) 2A or T8A/250V
Environmental Specications
• The BioProle FLEX2 ESM should be operated indoors. Keep the working area 
around the system free of dirt & debris, corrosive fumes, and excessive temperature 
changes.
• Ambient operating temperature between 15°C to 30°C (59°F to 86°F)
• Altitude up to 2500 meters (8202 Feet)
• Relative Humidity of 20 – 85% (without condensation)
• Installation Category (II)
• Pollution Degree (2)
Dimensions (ESM unit only)
• Height:  6.9 in. (17.53 cm) 
• Width:  9.7 in. (24.64 cm) 
• Depth:  18.2 in. (46.23 cm)
Weight:
• 8.6 lbs. (3.9 kg) Without Reagent Cartridge
• 17.8 lbs. (8.1 kg) With Reagent Cartridge