42
CD Indicators
CommandCommand
CommandCommand
Command
CharacterCharacter
CharacterCharacter
Character
DescriptionDescription
DescriptionDescription
Description
1=percent
2=parts counting
3=animal weighing
4= checkweigh
xAWxAW
xAWxAW
xAW Set Animal Level to "x", where x= 0 to 3.
0= minimum filtering
1=
2=
3= maximum filtering
xExE
xExE
xE Turn Auto restart in Animal mode on or off.
Where x=0 is off and x=1 is on.
xTxT
xTxT
xT Set tare weight. Uses real number "x" as current tare weight. Weight must be entered in
Calibration Units.
IDID
IDID
ID Print Current User ID String.
xIDxID
xIDxID
xID Program User ID String, 1-8 characters.
ACAC
ACAC
AC Abort Calibration.
xUCxUC
xUCxUC
xUC Start User Calibration. Uses integer number "x" as calibration weight.
LELE
LELE
LE Show last error code. Response: "Err: x.x".
xSxS
xSxS
xS Turn Print Stable Only on or off.
Where x=0 is off and x=1 is on.
TIMETIME
TIMETIME
TIME Print current time. Note, a "?" will follow if time has not been set.
mm/dd/yymm/dd/yy
mm/dd/yymm/dd/yy
mm/dd/yy Set date command and remove invalid date indicator.
SETDATESETDATE
SETDATESETDATE
SETDATE
hh:mm:sshh:mm:ss
hh:mm:sshh:mm:ss
hh:mm:ss Set time command and remove invalid time indicator. Time must be entered in 24 hr. format
SETTIMESETTIME
SETTIMESETTIME
SETTIME
DATEDATE
DATEDATE
DATE Print current date. Note, a "?" will follow if date has not been set.
SWSW
SWSW
SW Shows Lockswitch status.
ZZ
ZZ
Z Same effect as pressing ">0<" Key.
NGNG
NGNG
NG Same effect as pressing the "G/N/T" Key. Cycles through Net, Gross, and tare displays if
possible.
EscMEscM
EscMEscM
EscM To print out a list of the scale selections.
EE
EE
E Will start animal cycle when in animal weighing mode.
RS232 COMMAND TABLE
4.7 Printing Data (Cont.)
4.7.2 Output Formats (Cont.)