Defender 3000 Indicators EN-33
4.4 Readout Menu
Enter this menu to set user preferences. Default settings are in bold.
Note: / refers to not applicable in the table above.
Reset [reset]
Reset the readout menu to factory defaults.
NO = do not reset.
YES = reset
Stability [StAblE]
Set the amount reading can vary before the stability symbol turns off.
0.5d = 0.5 of a scale division
1d = 1 scale division
2d = 2 scale divisions
5d = 5 scale divisions
Zero [ZErO]
Set the percentage of scale capacity that can be zeroed.
2% = zero range is within +/-2%
100% = zero range is within +/-100%
Filter [FILtEr]
Set the amount of signal filtering.
LOwW = faster stabilization time with less stability.
mMEd = normal stabilization time with normal stability.
HIGH = slower stabilization time with more stability.
AZT [AZt]
Set the automatic zero tracking feature.
OFF = disabled
0.5d = the display will maintain zero until a change of 0.5 of a divisions per second has been
exceeded.
1d = the display will maintain zero until a change of 1 division per second has been exceeded.
3d = the display will maintain zero until a change of 3 divisions per second has been exceeded.
Backlight [B.LIGHt]
Set the display backlight feature.
OFF = backlight is disabled.